gossen luna pro calibration

All we really do with the adjustable lightbox is making the cell assume its characteristic resistances for a given set of illumination levels in sequence, repeatedly. However, some months ago I dropped the meter and it has been sticking, intermittently. Gossen Profisix / Lunasix F - Camera-wiki.org By By This camera manual library is for reference and historical purposes, all rights reserved. A 1N4148 should do. http://www.butkus.org/chinon/batt-adapt-us.pdf Please support this site with a Paypal donation by clicking the button above. I recommend you buy a meter that can be calibrated when it drifts out of calibration, as all meters do. info@gossen-photo.de, 2023 GOSSEN Foto- und Lichtmesstechnik GmbH. Both scales do give readings. Im thinking feed it from a 9 volt, tune it 2.7 volts, send to the contact with a dummy batter with appropriate contacts. A big advantage of your external exposure meter is the incident light measurement which prevents object-related incorrect exposures. Don't loose the meter zero adjusting screw. You recommend any small diode. For high light: use EV12 (19 on the dial) and EV8 (13) instead, and tweak the right two pots. #1. Equally, producing an accurate EV0 for calibrating low-light scale might turn tricky, so you can instead cheat, put the filter in (decoupled from its drive spring) and measure EV6 instead. Three of which I have recalibrated to take the Silver Oxide 1.55v batteries, and it would be tough for me to give up any of them. Gossen Luna Pro: readings way off - Photrio.com Photography Forums If so, best thing would be to test the CdS cell use a controlled setup ran against a known good meter and register resistance of the cell for every EV level in the range. I took apart my meter and, typical of my efforts, some parts fell out that I can't figure out how to put back. The upper half of the case will lift off. It takes me about 30 minutes.
, The two POTs on the left are for low light readings. We would like to give you a brief overview, which values are relevant to the assessment of horticulture lighting and how they can be measured. Choose the camera, the lens and possibly the film you work with most and define this as the standard setup for all further calibrations. Once set for the film you use, should not need setting again if you make a note of the adjustment details. UNFORTUNATELY, while I was screwing around the half-silvered mirror popped out. Remember, that each evenly lit, preferably white or gray surface you target, can provide you with two readings: one with and one without the ND filter. A black plate with the incident metering dome. Now if you mess up don't blame me, send me your unit and I will adjust it for you. There are two approaches to calibration: you can either do it directly, by exposing the meter to different light levels and succesively adjusting the trimm-pots, or indirectly, by recording the LDR characteristics and calibrating with substitute resistors. What you claim is a common belief, but it is easily proven wrong. Lina-Ammon-Str. For collectors, the English website www.myphotoweb.com by Ed van der Aa is a veritable treasure trove of the history of GOSSEN and information on the historic equipment. My next project will be restoring the old selenium meter of a Yashica-mat LM. or use LD6806 2,5V version? Light is detected using a CdS (Cadmium Sulphide) LDR (light dependent resistor) that directly influences amount of current allowed into the galvanometer. Marek, this is a wonderful article which should make calibrating my Lunasix 3 easy. One of the most important characteristics of a luxmeter is matching to the sensitivity of the human eye V(). How to calibrate a Gossen Luna Pro - Classic Manual Film Cameras Augie. Does the diode/SR-44 mod work on either one? A flat screwdriver with a narrow blade (2-3mm). Theres a neutral density filter of 6 stops attenuation installed in front of the LDR. An old flat type from a phone or camera (can be charged in original device) can be taped to the back of the meter. Anyone know how to Calibrate Gossen Luna Pro? | Rangefinderforum It is a good idea to make a chart
, with EV-3 and 2 Low.. Ev6 and 11 Hi for the low end, Ev8 and 13..Ev16 and 21 for the HI end.
, Adjust the low light to these readings and the high light to these readings. Sorry, but i made a mistake in the 1st formula. The styling, layout, and distinctive "computer" dial set the standard . The battery check was a bit high. Remove four: all that are outermost at the corners (in one corner there are two screws, remove only the one closer to the corner). Set the incident light metering mode on the external meter, with the white cap placed in front of the sensor, or the position on the probe activated. To calibrate your meter, we recommend the following method: You can convince yourself by continuing to take pictures of different subjects and to reassess them on the computer or after development on the light box. All I did was to put the lever arm through the hole in the corner, and swing the mirror part into its slot before closing the case. Flexible ac current probes (84 pages) Measuring Instruments Gossen MetraWatt METRACLIP 41 Operating Instructions Manual. Gossen Gossen Luna-Pro Analog Photography Light Meters for sale | eBay Your notes say that 0.6 0.99 is typical. Gossen Luna-Pro Photography Light Meters Shipping and returns View estimated shipping costs, delivery windows, and return policies at a glance. They should by about 6 stops, since the meter reads the same for both scales when its in. It makes it possible to precisely determine the amount of light at every place within the image. I shoot with so many cameras that I will leave a unit in each case so I don't have to look for one.
, Minolta is another good meter. Chat with Sales Dont overdo the heating thing, or you will deform the meter housing, or, worse, damage the movement inside. Remember you can cheat: you can set EV6 and put the ND in, when on low light range, to get EV0, and you can set EV6 without ND on high lights to get EV12. The same applies when changing the film material or post processing methode. Death is just mother natures way of telling you to slow down. How to calibrate a Gossen Luna Pro - Classic Manual Film Cameras It makes it possible to precisely determine the amount of light at every place within the image. Smooth as butter! For colour transparencies this gives a more saturated rendition and a result more pleasing than the correct exposure. However, later versions of the Gossen rang of meters such as the Lunalite use a PP3 9v battery that is readily available world-wide. YOu must go back and forth adjusting until you get it right. Now install the diode into your meter, reattach the battery wire, but DONT re-install the LDR yet. GOSSEN has implemented a special grid measurement function (GRID) in its new luxmeter series MAVOMASTER, MAVOPROBE for comfortable assessing workplace lighting. For calibrating using an ohmmeter you take the cell out anyway. For low light: set your lightbox to EV5 and adjust the slope pot for low-lights till the dial read `10. Free shipping on many items | Browse your favorite brands | affordable prices. I was thinking about using LD6806 low dropout voltage regulator instead of your solution. Also, not all meters come calibrated through the years. For a rough test: without ND in the way (on low-light scale) a reading off an average LCD monitor screen showing white should about just max out the scale (12). The 1977 Gossen Profisix (in US Luna-Pro sbc) is a professional "system" exposure meter that can be fitted with a range of electrically connected accessories offering additional measurement facilities, including flash, enlarger metering, and colour temperature. Thanks, that was a mistake: it should read EV14 (19 on the dial). I cant remember if you can overdrive the adjustment, but if you can, then turning it over a rotation would damage the movement. how to read and calibrate a Gossen Luna Pro SBC or profisix c______.3____1.6 At the present time I really don't want to be opening the case up and trying to make the calibration myself. While you can get these to match perfectly, it will throw the other end of scale off by much more. Gossen Luna Pro F meter: Battery check - Photrio.com Photography Forums I have a black Lunasix 3 (model 6D) which has hardly been used over the years. The meter draws a minute amount of power and so the voltage drop across a small-signal (thin and steep P-N junction) silicone diode will be less than 0.7V that a meter (testing at say 2mA) will show. Thanks for responding so fast, Today, Schottky diodes are a more common and reliable low barrier type that drop about 0.4V, such as type 1N5711. The meter switch on the right side opens a needle trap so the needle can swing freely. If you collect enough data, you can pick a few fixed resistors and put them in place of the LDR cell during calibration. You can also get one directly from Bogen/Gossen. Please some information about the meter. Simple and highly effective. The use of an exposure meter pays off in order to be able to keep the lighting situation under control at a photo shoot. ON Semiconductor NCP551SN27 (of their NCP551 series) drops about 0.3V so a regulated 2.7V will be maintained until your battery goes below 3V (solder a 100K resistor between Vin and Enable). Recommended Posts. Next step is to completely assemble the meter and check the calibration throughout the range. (Not working with EVs because Im calibrating this meter to work with my new sony fs100 video camera, which only offers ISOs of 3200,1600, and 500, which I have to traverse to match the Lunas sensitivity.) Thank you. thanks. I have a slightly different question you may know the answer to. Chat with Sales. First, disassemble the meter and unsolder the negative battery wire as before. Soldering iron, some solder and a little flux (or colophony). Toggling 2-3 times gets things working again. RE insert batteries. What would cause the meter needle to stick at the extremes of the scale? r 3.3 +, low range kymtmanDecember 8, 2008 in Classic Manual Film Cameras,

Remove Batteries, Zero meter. Nothing obvious, and no bad behavior while the meter was apart. Normal, center-weighted is best. It is two meters in one. Request Repair / Calibration 2nd: Adding a series resistor Rs = Rgalv (Rgalv || Rp) = Rgalv Rgalv^2 * 6.75 / (Rgalv * 7.75). Can you recommend a specific diode for the job, which did you use in the example posted above? If a gray card with 18% reflectance is used as the reference object and if it is measured format filling in reflected light measurement mode, we obtain the same exposure values as in the incident light measurement mode. V() is the spectral luminous efficiency of the human eye in daylight. 1 of 3 Next To calibrate your meter, we recommend the following method: Set the same ISO value, e.g. Those batteries are 1.5 volt alkaline The other gossen says to use two 1.35 volt mercury cells which are no longer manufactured. Because of this, in most cases it is possible to readjust the meter for 1.55V silver oxide cells with no further hardware modifications. ON insert batteries. correspond, and note it down. I had Lunasix 3 that suddenly stopped working and using you article was able to open it but I couldnt get any improvement, think the galvanometer faulty, so I bought a replacement. it's a good idea for removing th. Ive been working on your LDR calibration sheet. You dont need values every 1/3 stop, but try to avoid gaps exceeding 2 stops. Today I was working on my Luna Pro, and its been years since I last worked on them. If the light meter does not initially provide the values you expect, we recommend that you read this entry and perform the calibration described below. It won't fit there, though, without binding the meter switch. While you can revive these meters using silicone solar cells, the spectral response will differ and sensitivity curve will be shot. The aperture and shutter speed determined by the meter will be manually transferred to the camera and the first shot will be taken with these values. Model Reviewed: Luna-Pro F. Strength: the layout is extremely readable, and shows you EVERYTHING at once! Do you have a multimeter? I dont know, what internal differences are between my and yours models. questions, or feedback. Article about servicing the Gossen Luna Pro, Luna Pro S, Lunasix 3, and Lunasix 3s. Members; 621 Share; Posted May 12, 2003. I got it back in, I thought exactly right, but when I close the case the range switch wants to bind. Something like this: https://www.amazon.com/gp/product/B06XRN7NFQ/ref=ppx_yo_dt_b_asin_title_o05_s00?ie=UTF8&psc=1 ? Ive checked the voltage when the meter is on, and it looks correct. Download it to a computer with a calibrated monitor because this is the only way to ensure that you are correctly assessing the images. 22 . Independent of the matter of its source, it is an interesting question. Whether positive or negative, your feedback helps us continually improve the Tek.com experience. l_______.6___.3 If I adjust the slope pot further to get near 5, Thats what I did, and how I did it is described below, note, however, that this method requires some idea about basic electronics and may not actually be easier for you than the direct approach. I think it's ironic that the meter will give you a reading that you can't read unless you use a light to read it by!
,

How to calibrate a Gossen Luna Pro

,

Send it to Quality Light Metric in Los Angeles.

,

Thanks so much for this info and the pictures!
, I just came across this after posting elsewhere about an issue i have with a lunapro.
, It has a battery adapter kit for it. Since CdS cells will vary from piece to piece, I cant tell you what your cells resistance at EV10 or EV5 is, but it will be the same at EV10 each time. It was the company's first light meter not based on a selenium light receptor. high range With a national network of resources, a robust quality program and broad capabilities, we offer standard and custom solutions to deliver fast, accurate calibration when and how it's needed. What do you think about voltage 2,8 instead of 2,7? Free shipping. it is packed with info, with flash capabilities, and is easy to read. The Gossen Luna-Star F2 - Digital Incident, Reflected and Flash Light Meter apparently has been discontinued but is selling in the $250 range used. Otherwise the idea of an LDO crossed my mind, it was just not worth it You anyway will need to calibrate the whole business.